Scope of Calibration Services
Service Description Range CMC
  Pressure
Analog Pressure Gauge 0.0014 MPa to 110 MPa 0.01 %
Digital Pressure Gauge 0.0014 MPa to < 0.012 MPa *0.009 %
0.012 MPa to 7 MPa *0.008 %
> 7 MPa to 110 MPa *0.016 %
Digital Pressure Gauge 0.0014 MPa to < 0.012 MPa *0.009 %
(Absolute Mode) 0.012 MPa to 7 MPa *0.008 %
Pressure Transducer /
Transmitter with Electrical Output
0.0014 MPa to 70 MPa *0.015 %
Digital Vacuometer 0.0014 MPa to 0.101 MPa
(Absolute)
*0.011 %

  Electrical Signals
DC Current Source 10 μA to 100 μA 8 nA
> 100 μA to 1 mA 40 ppm + 8 nA
> 1 mA to 10 mA 40 ppm + 70 nA
> 10 mA to 100 mA 50 ppm + 1 μA
DC Current Measurement 10 μA to 100 μA 10 nA
> 100 μA to 1 mA 40 ppm + 10 nA
> 1 mA to 10 mA 40 ppm + 70 nA
> 10 mA to 100 mA 50 ppm + 1 μA
Resistive Decade or Resistance Simulator 1 Ω to 10Ω 20 ppm + 0.16 mΩ
> 10 Ω to 1 kΩ 20 ppm + 1.5 mΩ
> 1 kΩ to 10 kΩ 20 ppm + 17 mΩ
> 10 kΩ to 100 kΩ 20 ppm + 0.13 Ω
Resistance Measurement 1 Ω to 10 Ω 20 ppm + 0.16 mΩ
> 10 Ω to 1 kΩ 20 ppm + 1.6 mΩ
> 1 kΩ to 10 kΩ 20 ppm + 17 mΩ
DC Voltage Source 1 mV to 20 mV 0.8 μV
> 20 mV to 100 mV 20 ppm + 0.4 μV
> 100 mV to 1 V 10 ppm + 6 μV
> 1 V to 10 V 10 ppm + 30 μV
> 10 V to 100 V 10 ppm + 0.3 mV
DC Voltage Measurement 1 mV to 20 mV 0.8 μV
> 20 mV to 100 mV 20 ppm + 0.4 μV
> 100 mV to 1 V 10 ppm + 8 μV
> 1 V to 10 V 10 ppm + 30 μV
> 10 V to 60 V 10 ppm + 0.16 mV
> 60 V to 100 V 10 ppm + 0.29 mV

  Temperature
RTD Indicator or Controller -200 °C to 600 °C 0.03 °C
(Analog or Digital) > 600 °C to 800 °C 0.05 °C
Thermocouple Indicator or Controller -250 °C to < -100 °C 0.60 °C
(Analog or Digital) -100 °C to < 100 °C 0.07 °C
100 °C to 1000 °C 0.02 °C
> 1000 °C to 1370 °C 0.09 °C
> 1370 °C to 2100 °C 0.15 °C
> 2100 °C to 2300 °C 0.20 °C
RTD Simulator -200 °C to 600 °C 0.03 °C
> 600 °C to 800 °C 0.05 °C
Thermocouple Simulator -250 °C to -100 °C 0.60 °C
-100 °C to 700 °C 0.02 °C
> 700 °C to 1000 °C 0.04 °C
> 1000 °C to 1300 °C 0.06 °C
> 1300 °C to 2100 °C 0.09 °C
> 2100 °C to 2300 °C 0.20 °C
Temperature Transmitter for RTD -200 °C to 800 °C 0.05 °C
Temperature Transmitter for Thermocouple -250 °C to 2300 °C 0.10 °C
Thermometer with RTD Sensor -55 °C to -40 °C 0.20 °C
(Analog or Digital) > -40 °C to 0 °C 0.011 °C
0 °C (Ice Point) 0.015 °C
> 0 °C to 90 °C 0.010 °C
> 90 °C to 156 °C 0.015 °C
> 156 °C to 232 °C 0.016 °C
> 232 °C to 420 °C 0.019 °C
> 420 °C to 661 °C 0.029 °C
Thermometer with Thermocouple Sensor > 0 °C to 157 °C 0.061 °C
(Analog or Digital) > 157 °C to 232 °C 0.056 °C
> 232 °C to 420 °C 0.047 °C
> 420 °C to 660 °C 0.040 °C
> 661 °C to 900 °C 0.60 °C
> 900 °C to 1100 °C 0.75 °C
RTD Sensor -55 °C to -40 °C 0.20 °C
> -40 °C to 0 °C 0.011 °C
0 °C (Ice Point) 0.015 °C
0.010 °C (Triple Point of Water) ± 0.003 °C (k=2)
> 0.01 °C to 90 °C 0.010 °C
> 90 °C to 156 °C 0.015 °C
> 156 °C to 232 °C 0.016 °C
> 232 °C to 420 °C 0.019 °C
> 420 °C to 660.323 °C 0.027 °C
660.323 °C (Aluminum Fixed Point) 0.015 °C
Thermocouple -55 °C to 0 °C 0.20 °C
(Basic Metals) > 0 °C to 200 °C 0.10 °C
> 200 °C to 350 °C 0.20 °C
> 350 °C to 500 °C 0.50 °C
> 500 °C to 650 °C 0.60 °C
> 650 °C to 900 °C 0.60 °C
> 900 °C to 1100 °C 0.75 °C
Thermocouple > 0 °C to 157 °C 0.061 °C
(Noble Metals) > 157 °C to 232 °C 0.056 °C
> 232 °C to 420 °C 0.047 °C
> 420 °C to 660.323 °C 0.048 °C
> 660.323 °C 0.035 °C
> 660.323 °C to 900 °C 0.60 °C
> 900 °C to 1100 °C 0.75 °C
Temperature Transmitter with RTD Sensor - 55 °C to 420 °C 0.20 °C
> 420 °C to 650 °C 0.60 °C
Temperature Transmitter with - 55 °C to 500 °C 0.20 °C
Thermocouple Sensor > 500 °C to 1100 °C 0.65 °C
Dry Block Calibrator Parameter: Stability
> -55 °C to 180 °C 0.04 °C
> 180 °C to 420 °C 0.10 °C
Parameter: Uniformity
> -55 °C to 180 °C 0.04 °C
> 180 °C to 420 °C 0.10 °C
Parameter: Deviation of Control Temperature
> -55 °C to 180 °C 0.04 °C
> 180 °C to 420 °C 0.10 °C
Liquid Baths Parameter: Stability
> -80 °C to 400 °C 0.20 °C
> 400 °C to 660 °C 0.60 °C
> 660 °C to 1100 °C 1.5 °C
Parameter: Uniformity
> -80 °C to 400 °C 0.20 °C
> 400 °C to 660 °C 0.60 °C
> 660 °C to 1100 °C 1.5 °C
Parameter: Deviation of Control Temperature
> -80 °C to 400 °C 0.20 °C
> 400 °C to 660 °C 0.60 °C
> 660 °C to 1100 °C 1.5 °C
   Notes:
  • The calibration and measurement capability (CMC) is the lowest uncertainty the Laboratory is able to obtain. The Laboratory may declare in their calibration certificates measurement uncertainties greater than its CMC, due to the relative contributions of the properties or characteristics of the standard or instrument under calibration.
  • The CMC identified by an (*) does not include all contributions coming from the instrument or standard calibrated or measured device.
  • All the services are performed in permanent installations of Presys.
  • Calibration Laboratory accredited by Cgcre according to ABNT NBR ISO/IEC 17025 by the number 193.
  • Cgcre is Signatory of the ILAC Mutual Recognition Arrangement.
Click here to download.